This actuator is often used to adjust the backrest in multiple types of healthcare beds, where the Quick Release function can be essential in emergency situations. The damper helps staff retain the emergency lowering of the backrest while at the same time ensuring a quick, yet comfortable, lowering experience for the patient.
What to consider when testing the LA40 QR Damper in your healthcare application
- We recommend a minimum 30-minute cool-down period between activations
- The type and size of testing weights matter for the performance of the actuator
- The position of weights on the backrest matters for the performance of the actuator
- Using weights instead of a real person will yield different test results as the damper is designed for real life use